Uncertainty evaluation of oxygen content in electrolytic iron powder by pulse melting reduction-infrared absorption method |
MA Cheng-xin1,SHI Xiao-hua2 |
(1. National Center for Quality Supervision and Inspection of Diamond Tools, Ezhou 436000, China 2. Hubei Special Equipment Inspection Institute Ezhou Branch, Ezhou 436000, China) |
|
|
Abstract According to the procedure of measurement uncertainty in analysis and testing, and combined with the measurement and analysis principle of oxygen analyzer, the uncertainty of determining oxygen content in electrolytic iron powder by pulse melting reduction-infrared absorption method was evaluated. Through analysis of the measurement process, the establishment of mathematical model, each uncertainty component was calculated. Results show that the relative standard uncertainty caused by measurement repeatability is 0.002 360, the relative standard uncertainty caused by standard substances is 0.000 188, the relative standard uncertainty caused by instrument calibration is 0.000 584, the relative standard uncertainty caused by sample weighing is 0.004 350, and the relative standard uncertainty caused by instrument indication value is very small and negligible, thus obtaining an expanded uncertainty of 0.000 490%. The result of determination of oxygen content in electrolytic iron powder by pulse melting reduction-infrared absorption method is as follows: [WM=(0.048 800±0.000 490)%,][k=2.]It provides a theoretical basis for the accuracy of oxygen content determination of electrolytic iron powder.
|
Received: 17 March 2017
Published: 12 October 2018
|
|
|
|
[1]钱庆长, 史静, 王静.高频红外吸收法测定铜中氧的不确定度评定[J].铜业工程, 2016, 137(1):96-96
[2] JJF1059.1-2012. 测量不确定度评价与表示[S].
[3]张翼明, 任旭东, 郝茜等.惰气熔融-红外吸收法测定硫化钐中氧的不确定度评价[J].稀土, 2014, 35(4):89-89
[4]李岩, 郭恺.氧氮分析仪测定钢中氧、氮含量的不确定度评定研究[J].现代工业经济和信息化, 2015, 5(7):45-45
[5] GB/T 5158.4-2011, 金属粉末 还原法测定氧含量 第4部分:还原-提取法测定总氧量[S].
|
|
|
|